Our Ductless Fuming Chambers, Drying Cabinets and Workstations utilize in-unit filters that must be changed periodically to ensure that the unit is working properly, and that the operator is fully protected from potentially hazardous particulate, toxic fumes, and noxious odors.

These ductless units utilize one or more of the following filters:

Prefilters are made of electrostatically charged fiber and are designed to trap larger particulate. These filters should be replaced every 1 to 3 months for optimal operation (depending on usage).

HEPA Filters trap smaller particulate (99.997% efficiency at 0.3 microns) and should be replaced every 18 to 24 months for optimal operation (depending on usage).

Carbon Filters protect the operator from toxic chemical fumes, vapors, and noxious odors. These filters should be replaced every 18 to 24 months for optimal operation (depending on usage).

See below to match your ductless unit with the filter(s) necessary for safe, efficient operation.

DrySafe Drying Cabinets

All DrySafe units use the same Prefilters, HEPA Filters & Carbon Filters.

CyanoSafe Cyanoacrylate Fuming Chambers

mini CyanoSafe (No. CAS24)
The mini CyanoSafe uses only Hybrid Bonded Carbon filtration.

Benchtop & Freestanding CyanoSafe Units

These fuming chambers all use the same Prefilters and Hybrid Bonded Carbon Filters.

Latent Fingerprint Dusting Stations

DownDraft Dusting Stations use only HEPA filtration (different unit widths require unique filters).

  • Downdraft Dusting Station, 24" (No. DDS24)
    These units use No. ACFHEPA8
  • Downdraft Dusting Station, 36" (No. DDS36)
    These units use No. ACFHEPA18
  • Downdraft Dusting Station, 48" (No. DDS48)
    These units use No. ACFHEPA22

Our standard Latent Print Dusting Stations use both Prefilters and HEPA Filters (different unit widths require unique filters).

Basic Carbon Forensic Workstations

These units use Prefilters and Bonded Carbon Filters (different unit widths require unique filters).

  • Basic Carbon Forensic Workstation, 32" (No. AC632C)
    These units use No. ACFPRE and No. ACF100
  • Basic Carbon Forensic Workstation, 48" (No. AC648C)
    These units use No. ACFPRE4 (2 each) and No. ACF10018 (2 of each required for operation)

Basic HEPA Forensic Workstations

These units use Prefilters and HEPA Filters (different unit widths require unique filters).

Advanced Forensic Workstations

These units use Prefilters, HEPA Filters and Carbon Filters (different unit widths require unique filters).

PCR Workstations

Our PCR Workstations come as "Dead Air Boxes" with no filtration—only shortwave UV lighting, or with Prefilters & HEPA Filters as well as shortwave UV lighting. Filtration varies depending on unit width.

Identification Workstations

These units all use the same Prefilters, HEPA Filters and Bonded Carbon Filters.

NARKsafe™ Illicit Substance Workstations

These units all use multiple sets of the same Prefilters, HEPA Filters and Bonded Carbon Filters (# needed for operation varies by unit width)

NARKsafe™ Workstation, 36" (requires 2 each of the above for operation)

NARKsafe™ Workstation, 48" (requires 3 each of the above for operation)

NARKsafe™ Workstation, 72" (requires 4 each of the above for operation)